Specifications – HF scientific Micro100 Laboratory Turbidimeter User Manual

Page 7

Advertising
background image

Micro 100 (5/10)

Page

1

Rev. 3.3


SPECIFICATIONS:

MICRO 100

Conforms to specifications set forth in EPA method 180.1
(Nephelometric Method)

§

MICRO 100 IR

Conforms to specifications set forth in ISO 7027: Water
Quality - Determination of Turbidity.

Measurement Range

0-1000 NTU

Accuracy

†, ††

±2% of reading or ±0.01 NTU whichever is greater

Repeatability

††

< ±1% of reading or ± 0.01 NTU whichever is greater

Resolution

0.01NTU in the range 0.00 – 9.99 NTU
0.1NTU in the range 10.0 – 99.9 NTU
1 NTU in the range 100 – 1000 NTU

Response Time

Less than 6 seconds

Recorder Output

Uni-directional RS232 output

Power Supply

UL, CSA approved 12V DC Wall Mount

Miscellaneous
Specifications

1. Year 2000 Compliant
2. Built in Diagnostics
3. Three Year Battery Backup With no External Power

Operating Temperature Range

10°C – 40°C (50°F – 104°F)

Sample Temperature Range

0°C – 50°C (32°F – 122°F)

Dimensions

10.75” W x 10” L x 4.75” H (237mm W x 254mm L x 121mm H)

Shipping Weight

2.5 kg (5.5 lbs.)

Warranty

1 Year from date of shipment


§

The specifications found in EPA method 180.1 are essentially the same as the specifications set out in method 2130B of the Standard Methods for the

Examination of Water and Wastewater 19

th

edition and the specifications set out in ASTM Standard Method D1889-94. The MICRO 100 meets, or exceeds,

the specifications set forth in these methods.

Instrumental accuracy measured under controlled laboratory conditions at 25°C (77°F)

††

Both the accuracy and repeatability specifications for the MICRO 100 are valid only for measurement of static (non-flowing) samples.