MICRO 200 BW 0-10 & 0-100 NTU (05/05)
Rev. 3.4

By pressing in Normal Cal... menu. - 10 NTU Turbidity Only

In this menu you can offset the NTU value on the measurement screen to compensate for
variations between process NTU readings and grab sample readings measured on a known
good instrument. Offset adjustment is limited to ± 0.11NTU in 0.01 NTU increments. Use
the

keys to adjust the offset level to meet your requirements.

By pressing in Sensor Setup menu. - 10 NTU Turbidity Only

Press to perform a full scale calibration. In this menu a full scale calibration
of the instrument can be done using a 10 NTU standard. Simply put the 10 NTU standard
in the sensor and push the

key to perform the calibration.

3.2.3 Sensor Setup - Continued

To perform calibration on any cal point, move the cursor using the appropriate "F" key to
highlight the Hz column and then push one of the keys.

By pressing in Normal Cal... menu. - Turbidity

By pressing in Sensor Setup menu. - Backwash

Page 16

EXTENDED SETUP

10NTU TURBIDITY

FULL SCALE CAL...

VER:M200NTUBW-8A23E

FULL SCALE CAL

10NTU TURBIDITY

PUT STANDARD IN SENSOR

CAL TIME
PRESS UP TO CALIBRATE

6.9327NTU 6933 HZ

F1

F4

F3

F1

F4

NORMAL CAL

BACKWASH

CAL POINT 1

1

∅∅.∅%T

1

∅.∅∅∅HZ

9

∅.∅% Τ

9

∅∅∅HZ

NORMAL CAL

10NTU TURBIDITY

CAL POINT 1

∅.∅2∅∅NTU

237HZ

CAL POINT 2

4.

∅∅∅∅NTU

4114HZ

CAL POINT 3

1

∅.∅∅∅∅NTU

97737HZ

OFFSET LEVEL

.∅∅∅∅

∅∅∅∅

∅∅∅∅

∅∅∅∅

∅∅∅∅

.∅∅∅∅

∅∅∅∅

∅∅∅∅

∅∅∅∅

∅∅∅∅ NTU

∅HZ

NORMAL CAL

10NTU TURBIDITY

CAL POINT 1

∅.∅2∅∅NTU

237HZ

CAL POINT 2

4.

∅∅∅∅NTU

4114HZ

CAL POINT 3

1

∅.∅∅∅∅NTU

97737HZ

OFFSET LEVEL

.

∅4∅∅

∅∅

∅∅

∅∅

∅∅

.∅∅∅∅

∅∅∅∅

∅∅∅∅

∅∅∅∅

∅∅∅∅ NTU

∅HZ

Advertising
This manual is related to the following products: