0 overview, 1 microtol series – HF scientific MicroTOL Online Turbidimeter User Manual

Page 8

Advertising
background image

MICRO TOL (5/10)

2

Rev. 4.8




1.0 Overview

The MICRO TOL process turbidimeter allows for the measurement of the turbidity of
process water on-line. The White Light MICRO TOL has been designed to meet the
design criteria specified by the US EPA 180.1 on turbidity measurement. The infrared
MICRO TOL was designed to meet the design criteria specified in ISO 7027 and DIN
27027 for the measurement of the turbidity of a sample. Both models have long life lamps.

Some models have ultrasonic cleaning. Refer to section 8.2 for more information.

A pressure regulator on the incoming line is a standard on all Micro TOL instruments and
will reduce pressures up to 1380kPa (200 PSI) down to (104kPa) 15 PSI.

1.1 The Micro TOL Series

The Micro TOL series instruments have a wide variety of options available. Refer to the
table below to determine which factory installed options are available.

Catalog

No.

Description RS-485

Modbus

Backlight

Ultrasonic

Cleaning

Range

NTU

Flow

Alarm

20053 Micro

TOL2

WL

Standard

Standard Standard

N/A 0-1000 Option

20054 Micro

TOL2

IR

Standard

Standard

Standard N/A 0-1000 Option

20055 Micro

TOL3

WL

Standard

Standard Standard Standard 0-100 Option

20056 Micro

TOL3

IR

Standard

Standard

Standard Standard 0-100 Option

20063 Micro

TOL4

WL

Standard

Standard Standard Standard 0-1000 Option

20064 Micro

TOL4

IR

Standard

Standard

Standard Standard 0-1000 Option

Advertising