0 normal operation – HF scientific MicroTPW Field Portable Turbidimeters User Manual

Page 11

Advertising
background image

MicroTPI & MicroTPW (5/10)

Rev. 1.9

5

3.0 Normal Operation

This instrument allows for the measurement of the turbidity of a grab sample. The
turbidity is reported in Nephelometric Turbidity Units (NTU) and Formazin
Nephelometric Units (FNU). Readings above 1100 NTU are outside the range of this
instrument.

Note: Nephelometric turbidity units (NTU’s) are numerically equivalent to Formazin

nephelometric units (FNU’s) (See Glossary).

Below are some typical screens and a description of when you should see them.

When turning on the instrument or after a calibration, this screen indicates that the
instrument is ready to read:

The instrument will display moving dashes on the screen for 6-7 seconds prior to
displaying a value. This will occur prior to taking a reading & prior to indexing. No
keypad presses will operate during this time.


After taking a reading, a typical display:

If the sample that is being measured has a turbidity that is higher than that which the
instrument can measure, the instrument will indicate this as an over-range condition by
flashing 1100:

Advertising
This manual is related to the following products: