PROCEDURE

1.

Rinse a clean colorimeter tube (0967) with sample water. Fill to the 10 mL line with sample. This tube is the BLANK.

2.

Select setting 2 on “Select Wavelength” knob and press “30 Second Read” button.

3.

Insert tube into colorimeter chamber and adjust to 100%T with “Set Blank” knob. This is the 100%T blank.

4.

Remove BLANK from colorimeter and insert SAMPLE tube into colorimeter chamber. Record %T as soon as reading

stabilizes.

5.

Consult calibration chart to determine dissolved oxygen concentration in parts per million (ppm).

DC1600 DISSOLVED OXYGEN CALIBRATION CHART

%T

9

8

7

6

5

4

3

2

1

0

90

0.0

0.1

0.1

0.2

0.2

0.3

0.3

0.3

0.4

0.4

80

0.5

0.5

0.6

0.6

0.7

0.7

0.8

0.8

0.9

0.9

70

1.0

1.0

1.1

1.1

1.2

1.2

1.3

1.3

1.4

1.4

60

1.5

1.6

1.6

1.7

1.7

1.8

1.9

1.9

2.0

2.1

50

2.1

2.2

2.3

2.4

2.4

2.5

2.6

2.7

2.7

2.8

40

2.9

3.0

3.1

3.1

3.2

3.3

3.4

3.5

3.6

3.7

30

3.8

3.9

4.0

4.1

4.3

4.4

4.5

4.6

4.7

4.9

20

5.0

5.2

5.3

5.5

5.6

5.8

6.0

6.1

6.3

6.5

10

6.7

7.0

7.2

7.4

7.7

8.0

8.3

8.6

8.9

9.3

0

9.8

10.2

10.8

11.4

12.1

13.0