LaMotte DC1600 Colorimeter User Manual

Page 74

Advertising
background image

PROCEDURE

1.

Rinse a clean colorimeter tube (0967) with sample water. Fill to the 10 mL line with sample.

2.

Select setting 6 on “Select Wavelength” knob and press “30 Second Read” button.

3.

Insert tube into colorimeter chamber and adjust to 100%T with “Set Blank” knob. This is the 100%T blank.

4.

Remove tube from colorimeter. Use 1.0 mL pipet (0354) to add 1.0 mL of *Phosphate Acid Reagent (V-6282).

Cap and mix.

5.

Use the 0.1 g spoon (0699) to add one measure of *Phosphate Reducing Reagent (V-6283). Cap and shake until

powder dissolves. Wait 5 minutes for full color development. Solution will turn blue if phosphates are present.

6.

At end of 5 minute waiting period, press the “30 Second Read” button and insert tube into colorimeter chamber.

Record %T as soon as reading stabilizes.

7.

Consult calibration chart to determine the phosphate concentration in parts per million (ppm).

DC1600 PHOSPHATE ~ LOW RANGE CALIBRATION CHART

%T

9

8

7

6

5

4

3

2

1

0

90

0.01

0.03

0.05

0.07

0.10

0.12

0.14

0.17

0.19

0.21

80

0.24

0.26

0.29

0.31

0.34

0.36

0.39

0.41

0.44

0.47

70

0.49

0.52

0.55

0.58

0.61

0.63

0.66

0.69

0.72

0.75

60

0.79

0.82

0.85

0.88

0.92

0.95

0.98

1.02

1.06

1.09

50

1.13

1.17

1.20

1.24

1.28

1.32

1.37

1.41

1.45

1.50

40

1.54

1.59

1.64

1.68

1.73

1.79

1.84

1.89

1.95

2.00

30

2.06

2.12

2.19

2.25

2.32

2.39

2.46

2.53

2.61

2.69

20

2.78

2.86

2.95

3.05

10

0