C a l i b r a t e

0 0 . 5 0

q, Q, p

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

16. When the value on the display

matches the concentration of the
turbidity standard, press the

*IOK

to select Set.

C a l i b r a t e

0 0 . 5 0

* S e t

q

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

Or press

ˆ press *IOK to return

the meter to the default setting.

C a l i b r a t e

0 1 . 1 5

* D e f a u l t

p

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

17. Press

*IOK to proceed to Turbidity

analysis. Press

OFF to turn the meter

off or press

… to exit to a previous

menu or make another menu
selection.

Note: The meter will remember the last scanned blank reading. It is not necessary to
scan a blank each time the test is performed. To use the previous blank reading,
instead of scanning a new one, scroll to Scan Sample and proceed. For the most
accurate results, the meter should be blanked before each test and the same tube with
tube positioning ring should be used for the blank and the reacted sample.