LaMotte Tracer SAL Meter User Manual

Page 16

Advertising
background image

Electrode Cleaning Recommendations

Do not soak the electrode in the solutions for longer than the
recommended length of time. To do so may cause a reference
potential shift which will cause a degradation in performance or
failure. When cleaning the probe, do not scratch or damage the
platinized electrode surfaces.

Contaminant

Cleaning Solution

Procedure

Water soluble
substances

Deionized water

Soak or scrub with a soft brush.
Rinse thoroughly with DI
water and dry.

Grease and
oil

Warm water and
household detergant

Soak or scrub with a soft brush,
maximun 10 minutes. Rinse
thoroughly with DI water and
dry.

Heavy grease

Alcohol

Soak for a maximum of 5
minutes. Scrub with a soft
brush. Rinse thoroughly with
DI water and dry.

Lime and
hydroxide
coatings and oil

10% Acetic acid

Soak until coating dissolves,
maximum 5 minutes. Rinse
thoroughly with DI water and
dry.