LaMotte TDS 6 Conductivity User Manual

Page 26

Advertising
background image

Automatic Calibration (for CON 6)
The automatic calibration allows the meter to be quickly
calibrated to any of the four widely used conductivity
calibration standards. For a list of calibration standards
refer to Table 1.
In the manual calibration mode, customized conductivity
calibration standard can be used to calibrate this meter.
1. Enter Advanced Set-up.
2. Press

or

button until [ACAL] appears on the

LCD. Press ENTER button.

3. Press

or

button to select either [Yes] or [no].

4. Press [ENTER] button to select. The meter will

return to the menu, [ACAL].

5. Press

or

button to move to the next menu or press CAL to exit to

measurement mode.

Setting the TDS Factor (for TDS 6)
The concentration of salts dissolved in solution increases the conductivity of
that solution. This relationship varies from salt to salt and is roughly linear
over a given range for a given salt. The TDS conversion factor is the number
used by the meter to convert from conductivity to TDS.
To calculate the TDS conversion factor refers to
Appendix 2 – Calculating TDS Conversion Factor.
TDS conversion factors for various types of salts can also
be found in chemical reference books.
The TDS conversion factor can be set between 0.4 and
1.0. The meter default is 0.5.
1. Enter Advanced Set-up.
2. Press

or

button until [tdS] appears on the LCD.

Press ENTER button.

3. Press

or

button to select a value between 0.4 to

1.0.

4. Press ENTER button to select. The meter will take

return to the menu, [tdS].

5 .Press

or

button to move to the next menu or press CAL to exit to the

measurement mode.

HOLD

ENTER

Press

E

NTER

to

confirm

selection;

or

Press

C

AL

to

escape

without

confirming

selection

Press

ENTER

to

confirm

selection;

or

Press

CAL

to

e

scape

without

confirming

selection

HOLD

ENTER

26

Advertising
This manual is related to the following products: