And c, Conductivity at t, That is about 5 °c to 10 °c different from t – LaMotte TDS 6 Conductivity User Manual

Page 38: And t

Advertising
background image

2. Wait for 5 minutes. Note T

1

and C

T1

(conductivity at T

1

).

3. Condition the sample solution and probe to a temperature (T

2

) that is

about 5 °C to 10 °C different from T

1

, and note the conductivity reading

C

T2

.

NOTE: Record the results for future reference. Ideally T

1

and T

2

should

bracket the measurement temperature, and should not differ by more than
5 °C.

4. Calculate the temperature coefficient of the solution according to the

formula shown above.

5. Enter the calculated temperature coefficient into the meter. Refer to

Temperature Coefficient.

The calculated temperature coefficient will not be applied to all the meter
readings.

WARRANTY
This instrument, excluding the probe, is guaranteed to be free of defects in
material and workmanship for three years from the date of the original
purchase. The probe is guaranteed for six months from the date of the original
purchase. If, in the specified time periods, either the probe or meter is found to
be defective, it will be repaired or replaced without charge, except for
transportation expenses. This guarantee does not cover the batteries.
The guarantee is void under the following circumstances:

Operator negligence
Improper Application
Unauthorized servicing

Limits of Liability
Under no circumstances shall LaMotte Company be liable for loss of life,
profits, or other damages incurred through the use or misuse of their products.

Packaging and Delivery
Experienced personnel at LaMotte Company assure adequate protection
against normal hazards encountered during shipping. After the product leaves
the manufacturer, the transporter assures all responsibility for its safe delivery.
Damage claims must be filed immediately with the transporter to receive
compensation for damaged good

38

Advertising
This manual is related to the following products: