LaMotte SMART 3 Electronic Soil Lab User Manual
Page 27
SMART3 Soil 1.11
27
Press
10.
or
to scroll
to the concentration of the
prepared standard, 0.30 in
this example.
NOTE: A maximum
adjustment of 10% is possible.
If an adjustment of over 10%
is attempted, Overrange will
be displayed.
002 Aluminum
0.30
ppm
%T/Abs
Calibrate
12:00:00 001/500
Press
11.
ENTER
to select
Calibrate. Two menu choices
will be offered, set calibration
and factory setting.
002 Aluminum
0.30
ppm
Set Calibration
Factory Setting
12:00:00 001/500
Press
12.
ENTER
to select Set
Calibration and save the
calibration. Or press
to scroll to Factory Setting.
Press
ENTER
to select Factory
Setting to revert to the factory
calibration. The screen will
display Storing... for about
1 second and the test menu
will appear. The calibration
has now been standardized
and the meter can be used for
testing. The standardization
can be removed by repeating
the calibration and selecting
Factory Setting.
002 Aluminum
Scan Blank
Scan Sample
12:00:00 001/500
Testing