LaMotte SMART 3 Electronic Soil Lab User Manual

Page 27

Advertising
background image

SMART3 Soil 1.11

27

Press

10.

or

to scroll

to the concentration of the
prepared standard, 0.30 in
this example.
NOTE: A maximum
adjustment of 10% is possible.
If an adjustment of over 10%
is attempted, Overrange will
be displayed.

002 Aluminum

0.30

ppm

%T/Abs

Calibrate

12:00:00 001/500

Press

11.

ENTER

to select

Calibrate. Two menu choices
will be offered, set calibration
and factory setting.

002 Aluminum

0.30

ppm

Set Calibration

Factory Setting

12:00:00 001/500

Press

12.

ENTER

to select Set

Calibration and save the
calibration. Or press

to scroll to Factory Setting.
Press

ENTER

to select Factory

Setting to revert to the factory
calibration. The screen will
display Storing... for about
1 second and the test menu
will appear. The calibration
has now been standardized
and the meter can be used for
testing. The standardization
can be removed by repeating
the calibration and selecting
Factory Setting.

002 Aluminum

Scan Blank

Scan Sample

12:00:00 001/500

Testing