Procedure – LaMotte SMART 3 Electronic Soil Lab User Manual

Page 60

Advertising
background image

60

SMART3 Soil 1.11

PROCEDURE

I. DILUTION OF SOIL EXTRACT

Use the 30 mL graduated cylinder (0418) to measure 10 mL of the soil extract

and transfer it to a 50 mL beaker (0944). Add 10 mL of deionized water, mix and

neutralize according to the procedure on page 50.

II. TITRATION A, CALCIUM & MAGNESIUM

Carefully read the LaMotte Direct Reading Titrator Manual (1649) before

performing the titrations described below.

Fill the test tube (0778) to the 5 mL line with the soil extract from above.

1.

Dilute to the 10 mL line with deionized water.
Add 5 drops of Calcium Magnesium Inhibitor Reagent (3922).

2.

Wait 5 minutes.

3.

Use a pipet (0364) to add 5 drops of *Calcium & Magnesium Buffer (5126).

4.

Add 10 drops of *CM Indicator (6522WT).

5.

Fill the Direct Reading Titrator (0384) with the Standard EDTA Reagent

6.

(5254). Insert the tip of the Titrator into the center hole of the test tube cap.
While gently swirling the tube, slowly press the plunger to titrate until the

7.

color changes from red to blue.
Read the test result from the scale where the large ring on the Titrator meets

8.

the Titrator barrel. Multiply by 5.16. This is Titration Value A.

III. TITRATION B, CALCIUM

Fill the test tube (0778) to the 5 mL line with the diluted soil extract. Dilute to

1.

10 mL with deionized water.
Add 2 drops of *Inhibitor Solution (9258).

2.

Add 2 drops of *TEA Reagent (3921).

3.

Add 8 drops of *Sodium Hydroxide w/Metal Inhibitors (4259).

4.

Add one Calcium-Hardness Indicator Tablet (T-5250) to the test sample. Cap

5.

and shake to dissolve the tablet. A red color will develop.
Immediately titrate the sample. Fill the Direct Reading Titrator with Standard

6.

EDTA Reagent (5254). Insert the tip of the Titrator into the hole in the cap of

the test tube.
While gently shaking the tube, slowly press the plunger to titrate until the red

7.

color changes to a clear blue and does not revert to red upon standing 1-2

minutes.
Read the test result from the scale where the large ring on the Titrator meets

8.

the Titrator barrel. Multiply by 5.16. This is Titration Value B.

Test P

rocedures