LaMotte SMART Spectro Spectrophotometer User Manual

Page 40

Advertising
background image

When finished inputting standards scroll down
(in the NEW CALIBRATION menu) to the
CALCULATE

selection. Select CALCULATE.

NEW CALIBRATION

ENTER STD.7

ENTER STD.8

* CALCULATE

From the SELECT DEGREES menu scroll to and
select either 1 DEGREE THRU.0 or 1 DEGREE.

SELECT DEGREES

* 1 DEGREE THRU.0

1 DEGREE

NOTE: 1 DEGREE THRU.0 calculates the best straight line fit through
the data points and intersects with the origin at 0 ppm, 0 absorbance.
This is a classical Beers Law calibration. 1 DEGREE calculates the best
straight line fit but without forcing the line through the origin. The
minimum number of standards needed for a calibration is one for 1
DEGREE THRU.0

and two for 1 DEGREE.

Press

ENTER/Q button to select the curve fit

type. The display will show the constants for the
best fit line.

K0=0.0000 E+00

K1=1.0000 E+00

K2=0.0000 E+00

K3=0.0000 E+00

Press

EXIT button to escape back to SELECT

DEGREES

menu. Press

EXIT button again to

escape back to Input Data Numbers menu.
Press

EXIT button to escape back to 101 H20

menu.

101 H2O

* NAME THE TEST

SELECT WAVELENGTH

NEW CALIBRATION

FORMAT RESULT

40

SMART SPECTRO SPECTROPHOTOMETER 05.04