1. Install the console assembly into the wall opening (see Figure 3).

2. The unit should be mounted in such a way that allows for project specific drywall thickness (see Figure 5 for

reference, using a 5/8" [16 mm] drywall setup).

3. Ensure that the mounting location corresponds to the

on-site rough-in locations for medical gas and electrical

rough-ins.

4. Once the console assembly is in the correct position, attach

the console assembly to the studs using the holes on the

side and the frame provided with the unit. (see Figure

4) Use hardware that is compliant with applicable local

building codes (supplied by others).

5. The unit should be mounted in such a way that, once

drywall is installed, the frame will protrude 5/8" (16 mm)

out of the wall (see Figure 5 for reference, using a 5/8" [16

mm] drywall set up).

Figure 3

Figure 4

mounting holes

Figure 5

5/8"

[16 mm]

Support Bracket

Supplied by Amico

5/8"

[16 mm]

drywall

thickness

Building stud:

face to line up with

unit stud frame