Description of features – Amico Hydraulic Patient Transfer Stretcher (S-H-300) User Manual

Page 5

Advertising
background image

To raise a siderail:

a. Pull the siderail up until it locks into place. A “click” will be heard as

the rail lock engages.

b. To make sure the siderail is fully latched, try pivoting the siderail

down.

To lower a siderail:

a. Grasp the top of the siderail and pull it slightly towards the foot of

the stretcher.

b. Locate and depress the siderail lever.
c. Rotate the siderail down while the siderail lever is depressed.

IV Pole Mounting Sockets

An IV pole mounting socket is located at each corner of the stretcher.

Siderail Positions

!

WARNING:

Evaluate patients for entrapment risk according to facility protocol and monitor patients appropriately. Ensure that all

siderails are fully latched when in the raised position. Failure to do either of these could result in serious injury or death.

NOTE:

Siderails are intended to be a reminder to the patient of the unit’s edges, not a patient-restraining device. When

appropriate, Amico recommends that medical personnel determine the proper methods necessary to ensure a patient

remains safely in the stretcher.

Mattresses

!

WARNING: Only mattresses that meet Amico specifications and recommendations should be used. Failure to do

so could reduce the effectiveness of the safety features and systems incorporated into the stretcher. Personal injury or

equipment damage could occur.

Push Handles

The push handles located at the head and foot ends of the stretcher are used to

move the stretcher.

Lift up the push handle and rotate it into the push position. The handle should

automatically lock into place.

Lower the push handle by pulling up on the locking pin and rotating it to the

stored position.

Brake/Neutral

The Brake/Steer pedals are located at both the head and foot of the S-H-

TRAUMA and it is equipped with a central brake system.

!

WARNING:

Unless transporting the patient, always set the brakes when the stretcher is

occupied. Reconfirm that the brakes are set before any patient transfer. Failure

to do so may result in injury or equipment damage.

To set the brakes, press one of the back brake pedals. All four casters, will lock.

To release the brakes, press the pedal in the neutral position.

Steering System

The Steering System is optional on the S-H-TRAUMA stretcher. The steer

function is used to keep the stretcher moving in a straight line and to make it

easier to turn corners.

Description of Features

Description of Features

8

Amico Beds

www.amico.com

9