DUAL LINE REGULATOR

On every Amico Automatic Gas Manifold the line regulator is capable

of maintaining a constant dynamic delivery of pressure at the

maximum designed flow rate of the system. For Oxygen, Nitrous

Oxide and Carbon Dioxide service, the line regulators are factory preset

at 55psi and for Nitrogen at 170psi.

LINE REGULATOR BALL VALVE

Quarter-turn ball valves are provided upstream and downstream of each

line regulator. These valves allow for removal and servicing of one of the

line regulators while the other is in use. All four ball valves are normally in

the open position.

Description of Parts

Pressure relief valves are installed downstream of all pressure regulators
and are set at no more than 50% above the setting of the pressure
regulator. Relief valves are capable of fully relieving the pressure at
the set point. All pressure relief valves in the manifold have piping
connections to allow for connection of a vent line to the outside of
the facility.

PRESSURE RELIEF VALVE

FRONT PANEL INDICATORS

Six front panel indicators monitor the status of the manifold.