Calibrating the temperature sensor – B&G Network Speed User Manual

Page 20

Advertising
background image

CALIBRATING THE TEMPERATURE SENSOR


Temperature calibration should only be carried out when the
built-in seawater temperature sensor (part of the speed
sensor) is considered to be inaccurate. The temperature of
the water will have to be measured with an accurate
temperature-sensing device; this value can then be entered
into the Network SPEED unit.

In the following example the unit is displaying a reading of
10.4

0

C, the water temperature is known to be 11.4

0

C. The

correct value is entered into the Network SPEED unit.



Press TEMP

key until
TEMP

0

C is

displayed.

Press SETUP
key twice to
display TEMP
CAL
.

Press ENTER
if to change.
The display
will flash.

Press ENTER
key to
memorise the
new units.

Press TEMP
key to display
sea
temperature.

Use ¿ or À
keys to
change the
displayed
units.