Bio-Rad Aurum™ Total RNA 96 Kit User Manual
Page 19
15
Section 9
Troubleshooting Guide
Problem
Possible Cause
Recommended Solution
Difficulty achieving Purge step in protocol
If sealing of wells with
–17 to –23 inHg
sealing tape restores
negative pressure
negative pressure, no
corrective action is
required; however, do
not leave sealed
Open vacuum regulator
Turn control knob fully
clockwise to close
regulator
Plate not seating on
Press down gently on
vacuum manifold
RNA binding plate to
seat
Manifold top not seated
Ensure that manifold
properly on base
top is seated properly on
base
Residue on gasket
Rinse manifold and
gasket with distilled
water and wipe dry
Gasket detached from
Press down to
manifold
reestablish uniform
adhesive contact or
replace gasket
Worn manifold gasket
Replace gasket
Cross-contamination Abrupt application of
Increase negative
between adjacent
negative pressure during pressure gradually
wells
elution
over 5–10 sec using
the vacuum regulator
Residual wash buffer on
Blot RNA binding plate
drip directors
drip directors with paper
towels; ensure that the
plate underside is dry