Bio-Rad Aurum™ Total RNA 96 Kit User Manual

Page 19

Advertising
background image

15

Section 9

Troubleshooting Guide

Problem

Possible Cause

Recommended Solution

Difficulty achieving Purge step in protocol

If sealing of wells with

–17 to –23 inHg

sealing tape restores

negative pressure

negative pressure, no

corrective action is

required; however, do

not leave sealed

Open vacuum regulator

Turn control knob fully

clockwise to close

regulator

Plate not seating on

Press down gently on

vacuum manifold

RNA binding plate to

seat

Manifold top not seated

Ensure that manifold

properly on base

top is seated properly on

base

Residue on gasket

Rinse manifold and

gasket with distilled

water and wipe dry

Gasket detached from

Press down to

manifold

reestablish uniform

adhesive contact or

replace gasket

Worn manifold gasket

Replace gasket

Cross-contamination Abrupt application of

Increase negative

between adjacent

negative pressure during pressure gradually

wells

elution

over 5–10 sec using

the vacuum regulator

Residual wash buffer on

Blot RNA binding plate

drip directors

drip directors with paper

towels; ensure that the

plate underside is dry