Model 7i Digital Force / Torque Indicator

User’s Guide


15

Parameter Description

Initial Delay

The time delay, in seconds, before the averaging sequence commences.

Averaging Time The time duration, in seconds, of the averaging sequence.
Trigger Load

The minimum load required to start the averaging sequence. Toggle between
measuring directions by pressing the DIRECTION key. Initial delay follows the trigger
load.


Upon completion of an averaging sequence, several functions can occur, as configured in the Auto
Settings
section:

Parameter Description

RS232/USB Output

Automatically output the average via RS-232 or USB at the completion of the
averaging sequence.

Mitutoyo Output

Automatically output the average via Mitutoyo (Digimatic) at the completion of the
averaging sequence.

Memory Storage

Automatically store the average at the completion of the averaging sequence.

Auto Zero

Automatically zero the reading at the completion of the averaging sequence. This
re-arms the indicator for a new averaging sequence as if ZERO has been
pressed.

Auto Zero Delay

The amount of time before the auto-zero occurs, in seconds:
Available settings:

1-10 sec., in 1 sec. increments, and 10-60 sec., in 5 sec.

increments.

Output Pin

Automatically toggle the SP1, SP2, or SP3 pins (active low). If not required, select
“NONE”. To stop a Mark-10 motorized test stand when averaging is completed,
specify SP2 if moving in the tension/counter-clockwise direction, or SP1 for the
compression/clockwise direction.
Note: Because the same pins are also used for set points, the stand will stop at
whichever condition occurs first.


After the parameters have been configured and the menu has been exited, press MODE until AVG is
displayed. Then press ZERO. Average mode is now armed, and the averaging sequence will commence
when the trigger load has occurred. The current status of the averaging sequence is displayed below the
primary reading, as follows:

Step Status

Description

1

TRIG WAIT

The trigger load has not yet occurred.

2

INIT DLY

The initial delay is currently taking place.

3

AVERAGING

The indicator is collecting readings. The status will be flashing until averaging has
been completed.

4

AVRG DONE Averaging has been completed. The average load is displayed in the primary

reading.


At the completion of the averaging sequence, the peak values are retained until ZERO is pressed.
Another averaging sequence may be started after ZERO has been pressed (or auto-zeroed). To exit
Average mode, press MODE and select the desired measuring mode.

AUTO SETTINGS

*

RS232/USB Output

* Mitutoyo Output
* Memory Storage
* Auto Zero
Auto Zero Delay

5 sec.