Section VIII — Battery Replacement

The Genius 2 Tympanic Thermometer and Base batteries should be replaced when the low battery icon is shown on the LCD

display. After the low battery icon is displayed, the user will be able to take approximately 100 temperatures before the dead

battery icon is displayed and the unit will not allow a temperature to be taken. To replace the batteries, access the battery

compartment by unscrewing the battery door that retains the battery cover. Note the polarity of the installed batteries. Remove

the old batteries and install fresh batteries, ensuring the correct polarity. Reinstall the battery cover and secure it with the screw.

Section IX — Mounting Instructions

If using the wall or cart mount options (sold separately) for the base unit, consult the installation sheet provided with the

respective system.

Section X — Troubleshooting

If the Genius 2 Tympanic Thermometer and Base is not functioning properly, check the following items:
Symptom

Action

Temperature reading unusually high

Check the probe cover for tears or gaps.

Temperature reading unusually low

Check the probe cover and thermometer tip for obstructions.

Check the patient ear canal for obstructions.

Low battery indicator lit

Replace battery

Dead battery indicator lit

Replace battery

Display blank

Replace battery

System error displayed

If system error “4”, then let the thermometer equilibrate in the

room for 20 minutes before using. For all other system errors, reset

the thermometer by installing a probe cover. If the system error

does not clear, send the thermometer to the facility Biomedical

Engineering department or contact the factory for available

service options. The service information is located in Section XII,

Customer Service.

The Genius 2 Tympanic Thermometer and Base advisory and alarm conditions are described in the Thermometer Display Icons

and Alarms subheading of Section V, Instructions for Use.

Section XI — Specifications

Clinical accuracy characteristics and procedures are available from the manufacturer on request. To verify accuracy, use a certified

blackbody as specified in EN 12470-5-2003, Annex C or use a Genius 2 Checker/Calibrator – order part number 303097.

Calibrated Accuracy Limits:

Ambient Temperature

Target Temperature

Accuracy

25°C

36.7°C to 38.9°C

± 0.1°C

(77°F)

(98.1°F to 102°F)

(± 0.2°F)

16°C to 33°C

33°C to 42°C

± 0.2°C

(60.8°F to 91.4°F)

(91.4°F to 107.6°F)

(± 0.4°F)