COVIDIEN Genius™ 2 Tympanic Thermometer User Manual

Page 13

Advertising
background image

English

Tympanic Thermometer and Base

10

Table of Contents

Calibrated Accuracy Limits (after Recalibration*):

Ambient Temperature

Target Temperature

Accuracy

16°C to 33°C

36°C to 39°C

± 0.2°C

(60.8°F to 91.4°F)

(96.8°F to 102.2°F)

(± 0.4°F)

16°C to 33°C

less than 36°C

± 0.3°C

(60.8°F to 91.4°F)

(less than 96.8°F)

(± 0.5°F)

16°C to 33°C

greater than 39°C

± 0.3°C

(60.8°F to 91.4°F)

(greater than 102.2°F)

(± 0.5°F)

*Post recalibration accuracy using the Genius 2 Checker/Calibrator may not necessarily be equivalent to factory calibration.

Displayed Temperature Measurement Range:

Temperature Range depends on Site Mode as follows:
Mode

Range °C

Range °F

Ear

33.0 to 42.0

91.4 to 107.6

Oral

33.6 to 42.0

92.5 to 107.6

Core

34.0 to 42.0

93.2 to 107.6

Rectal

34.2 to 42.0

93.6 to 107.6

Ambient Temperature Range:

16°C to 33°C (60.8°F to 91.4°F) where as CEN defines the standard as 16°C to 35°C (60.8°F to 95°F), 10 to 95 %RH, non-condensing.

Storage Temperature Range:

-25°C to 55°C (-13°F to 131°F), up to 95% RH non-condensing. If the unit is stored at extremes, it is recommended that the unit be

checked on the field calibration checker or at the factory before returning to service.

Clinical Repeatability:

Meets Section A.5 of EN 12470-5: 2003 (E) per Covidien llc technical report. Data is available from Covidien llc on request.

Response Time:

Less than 2 seconds

Pulse Timer:

60 seconds

Temperature Resolution:

0.1°C or 0.1°F

Power:

3 AAA alkaline batteries

Battery Life:

Minimum of 15,000 temperature readings

Size:

Thermometer — 17.8 cm (7”)

Base — 20.3 cm (8”)

Weight:

Thermometer (with batteries) — 160 grams

Base — 100 grams