Troubleshooting, Xi. specifications, X. troubleshooting – COVIDIEN Filac™ 3000 AD/ADA Electronic Thermometer User Manual
Page 13
Filac
™
3000 AD/ADA Electronic Thermometer
10
X. Troubleshooting
1. If device fails to function properly after washing & reconnecting probe, rinse all contacts in de-ionized water and blow dry.
2. Icons indicate all other possible errors and remedies.
3. Device Error Codes:
E01 - System error during synchronization
E02 - System error during patient thermistor calibration
E03 - System error during heater thermistor calibration
E04 - System timing error
E05 - Heater error
E06 - LCD communication error
E07 - Shut off error
P01 - Probe configuration (or no probe connected) error
P02 - Direct Mode patient thermistor unstable or out of range
P03 - Direct Mode heater thermistor unstable or out of range
P04 - Predict Mode patient thermistor unstable or out of range
P05 - Predict Mode heater thermistor unstable or out of range
P06 - Unable to pre-heat probe tip
4. No user-serviceable parts inside.
5. Refer to a factory or qualified Biomedical Technician for service.
XI. Specifications
Medical Electrical Equipment
Filac 3000 AD/ADA Electronic Thermometer
Classified with respect to electrical shock, fire, and mechanical hazards only in accordance with UL60601-1
Classified with respect to electrical shock, fire, and mechanical hazards only in accordance with CAN/CSA C22.2 No. 60601-1-08
Dimensions (Approximate, without base):
in.- 3.2 W x 6.0 H x 2.1 D ; mm - 84 W x 152 H x 56 D
Dimensions (Approximate, with base):
in.- 4.1 W x 7.5 H x 3.1 D ; mm - 105 W x 190 H x 79 D
Weight (Approximate, without base):
13.5 Ounces; 385 Grams
Temperature Measurement Range:
30°C to 43°C; 86°F to 109°F
Ambient Operating Temperature Range:
10°C to 40°C; 50°F to 104°F; at 10% - 95% RH, non-condensing.
Average prediction times
(after insertion into measurement site):
Oral (Quick Mode):
3-5 seconds (non-fever temps),
8-10 seconds (fever temps)
Oral (Standard Mode):
6-10 seconds
Axillary Mode:
8-12 seconds
Rectal Mode:
10-14 seconds
Direct Mode (All Sites):
60-120 seconds
Typical times for switch to Direct Mode:
If no measurement site detected: 60 seconds
(after pulling probe from probe well)
If temperature does not stabilize: 70 seconds
(after measurement site insertion)
Pulse Timer:
60 Second count with a “beep” at 15 seconds, 2 “beeps” at 30 seconds, 1 “beep” at
45 seconds and 2 “beeps” at 60 seconds