Troubleshooting, Xi. specifications, X. troubleshooting – COVIDIEN Filac™ 3000 AD/ADA Electronic Thermometer User Manual

Page 13

Advertising
background image

Filac

3000 AD/ADA Electronic Thermometer

10

Table of Contents

X. Troubleshooting

1. If device fails to function properly after washing & reconnecting probe, rinse all contacts in de-ionized water and blow dry.
2. Icons indicate all other possible errors and remedies.
3. Device Error Codes:

E01 - System error during synchronization

E02 - System error during patient thermistor calibration

E03 - System error during heater thermistor calibration

E04 - System timing error

E05 - Heater error

E06 - LCD communication error

E07 - Shut off error

P01 - Probe configuration (or no probe connected) error

P02 - Direct Mode patient thermistor unstable or out of range

P03 - Direct Mode heater thermistor unstable or out of range

P04 - Predict Mode patient thermistor unstable or out of range

P05 - Predict Mode heater thermistor unstable or out of range

P06 - Unable to pre-heat probe tip

4. No user-serviceable parts inside.
5. Refer to a factory or qualified Biomedical Technician for service.

XI. Specifications

Medical Electrical Equipment

Filac 3000 AD/ADA Electronic Thermometer

Classified with respect to electrical shock, fire, and mechanical hazards only in accordance with UL60601-1

Classified with respect to electrical shock, fire, and mechanical hazards only in accordance with CAN/CSA C22.2 No. 60601-1-08

Dimensions (Approximate, without base):

in.- 3.2 W x 6.0 H x 2.1 D ; mm - 84 W x 152 H x 56 D

Dimensions (Approximate, with base):

in.- 4.1 W x 7.5 H x 3.1 D ; mm - 105 W x 190 H x 79 D

Weight (Approximate, without base):

13.5 Ounces; 385 Grams

Temperature Measurement Range:

30°C to 43°C; 86°F to 109°F

Ambient Operating Temperature Range:

10°C to 40°C; 50°F to 104°F; at 10% - 95% RH, non-condensing.

Average prediction times
(after insertion into measurement site):

Oral (Quick Mode):

3-5 seconds (non-fever temps),
8-10 seconds (fever temps)

Oral (Standard Mode):

6-10 seconds

Axillary Mode:

8-12 seconds

Rectal Mode:

10-14 seconds

Direct Mode (All Sites):

60-120 seconds

Typical times for switch to Direct Mode:

If no measurement site detected: 60 seconds

(after pulling probe from probe well)

If temperature does not stabilize: 70 seconds

(after measurement site insertion)

Pulse Timer:

60 Second count with a “beep” at 15 seconds, 2 “beeps” at 30 seconds, 1 “beep” at
45 seconds and 2 “beeps” at 60 seconds