Humboldt HM-4165 3-Cell Manual Control Panel User Manual

Page 5

Advertising
background image

3

which connects to the volume change device. This adapter is supplied

with the panel.

TRIAXIAL CELL CONNECTION
The triaxial cell has compression fittings on the front as shown in

Figure 4. The cell connection on the triaxial cell connects to the water

connection at the Cell connection on the front of the panel. The base

connection on the triaxial cell connects to the volume change device as

shown in Figure 2 and not directly to the panel.

DE-AIRING TANK CONNECTION
The de-airing tank has connections for the tap water in, de-aerated

water out, and air/vacuum lines. These lines connect to the back side

of the panel at the corresponding push-to-connect fittings as shown in

Figure 2.

ELECTRICAL CONNECTION
The only electrical connection required is to connect the AC to DC

transformer into the back of the digital display and then plug the

transformer into a 110/220 VAC electrical socket as in Figure 5.

OPERATION
After making all of the necessary connections, check to make sure all

of the valves on the front of the panel are set to the off position. Also,

loosen the jam nuts on the regulators as shown in Figure 3.

AIR PRESSURE ADJUSTMENT
There are a total of seven regulators on the panel. The main regulator,

located below the gauge labeled SYSTEM AIR PRESSURE, controls the

supply of air for all of the other regulators. There are six regulators on

the front of the panel near the center which control the output pressure

to the bladder cylinders for cell pressure and back pressure control.

There is also a small regulator at the front of the panel in the upper

right corner labeled AUXILIARY AIR PRESSURE. This regulator is used to

control up to 10 psi of pressure to assist in the purging of water either

from the triaxial cell or from the de-airing tank. This regulator has a

locking knob which must be pulled out to adjust the pressure or pushed

in to lock it into position.