STERILIZATION INSTRUCTIONS

• Miltex recommends that the sterilization trays be processed according to the sterilization wrap

manufacturer’s instructions prior to sterilization to maintain sterility of internal components/items

and for proper aseptic presentation to the surgical/dental field.

• For Prevacuum Sterilizer:

o

Wrapped trays and instruments should be exposed to 132° C to 135° C (270° F to 275° F)

for at least four minutes.

o

Dry 20 to 40 minutes.

• For Gravity Displacement Sterilizer

o

Wrapped trays and instruments should be exposed to 132 C to 135 C (270 F to 275 F) for at

least 30 minutes, or 121° C to 123° C (250° F to 254° F) for at least 55 minutes.

o

Dry for 20 to 50 minutes.

• Variables that may affect drying times include: loading density of the tray, instrument configuration,

total contents of the sterilizer, steam quality, equipment maintenance and others.

MAINTENANCE OF TRAYS

• Always inspect for cleanliness or damage before use.

• Make sure all latches and handles are secure and in working order.

SYMBOLS USED ON LABELING

Manufacturer

Authorized Representative in the European Community

Catalog number

Lot number

Non-sterile: Sterilize prior to use

CAUTION: See warnings or precautions

Consult instructions for use

Product complies with requirements of directive 93/42/EEC for European Union Class 1

medical devices

CAUTION: Federal (USA) law restricts this device to sale by or on the order of a

dentist.

Miltex, Inc.

Miltex GmbH

589 Davies Drive

78604 Rietheim-Weilheim

York, PA 17402

Germany

800-645-8000

©2008 Miltex, Inc. All rights reserved

.

www.miltex.com

SH-030000 Rev.1

10/08

Advertising