Unpacking and installation – Iris Sample Processing StatSpin® CytoFuge 2 Compact Cytocentrifuge System User Manual

Page 5

Advertising
background image

1

Section

1

Unpacking and Installation

Inspect Packaging

The CytoFuge 2 and its accessories are delivered in one carton. If the cytocentrifuge or accessories
have suffered any damage in transport, please inform your carrier immediately

.

NOTE: Save shipping carton and components to simplify return should service be required.

Verify Contents

The package contains:
One CytoFuge 2 cytocentrifuge (Model No. M801-22)
One Universal Switching Power Supply
(Product No. 01-3553-001, APS Model No. AD-744-1240)
One grounded line cord (for North American use only)
One Operator’s Manual
One Accessory Carton
The Accessory Carton consists of:

Product Number

3-well Cell Concentrators and 3-well gaskets, package of 4 each

CC03

1-well Cell Concentrators and 1-well gaskets, package of 4 each

CC01

Stainless Steel Clips for Cell Concentrators, package of 8 each

CLIP

Backing Plates for Cell Concentrators, package of 4 each

BP01

Filter Concentrators, 2 bags of 4 each

FF01

Stainless Steel Clips for Filter Concentrators, package of 4 each

FFCL

Install System

1. Place the CytoFuge 2 on a level surface suitable for laboratory instrumentation.
2. Maintain a 300mm clearance boundary around the centrifuge for ventilation and safety.
3. Position the CytoFuge 2 away from direct sunlight and sources of heat or cold.

(See Appendix A for specifications.)

Connect Power

Plug the power supply into a grounded outlet supplying the voltage and frequency indicated on the
power supply. When power is connected, two “beeps” will sound and the cover lock will release. To
turn unit off completely, disconnect power located at the rear of the instrument.

WARNING - North American Installation:

Only use the power supply included with the unit.

Use of other power supplies or transformers will damage the CytoFuge 2 electronics and
void the warranty.

WARNING - Outside of North America:

Do Not Use the Line Cord Supplied. Use power cord

for at least 1.0 Amp or more with an IEC320/CEE22 female connector and male connector
suitable for the power outlet to be used.