Operating instructions – Iris Sample Processing StatSpin® CytoFuge 2 Compact Cytocentrifuge System User Manual

Page 9

Advertising
background image

5

Section

4

Operating Instructions

Quick Start Instructions

Details regarding each instruction can be found in the following sections.

1.

Assemble the number and type of concentrators required.

2.

Install the concentrators in the rotor in a balanced arrangement.

3.

Load the concentrators with sample. (Empty concentrators may be used for balance.)

4.

Screw on the rotor lid.

5.

Close and latch cover.

6.

Select time and speed.

7.

Press start.

When the timed cycle is complete, the rotor stops; three beeps sound; and the
interlock mechanism releases. The cover latch can then be squeezed to open.

8.

Unscrew the rotor lid.

9.

Remove concentrators.

10.

Aspirate fluid. (Cell Concentrators only)

11.

Disassemble concentrators.

12.

Recover slides for further processing.

NOTE: The CytoFuge 2 has no “on-off” switch, and therefore is normally left plugged in and “on”.

Detailed Instructions

Opening and Closing the Cover

The system’s electrically operated cover interlock mechanism prevents operation until the cover is
completely closed and latched, and prevents the cover from being opened while the rotor is turning.
When the cover is completely closed, it is locked and an operating cycle can be initiated.

The system is also equipped with a manually operated latch that holds the cover down after spinning
is complete. The interlock is automatically released at the end of the operating cycle or by pushing
the stop/open button. Squeeze the black latch pieces together to open the cover.

Cover Interlock By-pass

The electronically operated cover interlock mechanism can be released manually by inserting the
straightened end of a large paper clip or similar object into the small hole between the two buttons in
the front of the latch cover (facing the instrument). Manually push the lock lever inward about one
inch (25mm) to release the interlock mechanism if the stop/open button does not release the cover.

CAUTION - The cover interlock bypass is designed for emergency use only. If the
equipment is not used properly, safety may be compromised.