ITC AVOXimeter 4000 Operator’s Manual

Calibration

The ITC AVOXimeter 4000 is factory-calibrated and employs highly stable state-of-the-art light

e

Accura
code (see

cuvette calibration code whenever using a different

t number of cuvettes.

Note:

for each

Entering a Different Cuvette Calibration Code:

“Calibration” menu (a submenu of

sources.

Cuvett

cy

Calibration Code

of total hemoglobin measurements depends on using the correct cuvette calibration
page 1). The user must check the

lo

The lot number and calibration code are included on the carton label and bag
lot of cuvettes.

,

1.

Display the

see page 7).

2.

Press

followed by

to display the cuvette calibration code screen:

3.

Press

followed by

to enter a new value. A screen for entering a new

cuvette calibration code is displayed:

4.

Enter the new calibration code, using the number keys, then press

. A

confirmation

5.

Press

screen is displayed.

followed by

to confirm the new calibration code. The “Calibration”

menu is again displayed.

6.

Press

to return to the previous menu, if desired.

Re-Calibration

If the ITC AVOXimeter 4000 results are out of range with either blood samples or controls and
troubleshooting does not resolve the issue, re-calibration may be necessary. Please contact ITC
Technical Support for more information on re-calibration.