Interface and output, Maintenance and cleaning – Scilogex OS20-Pro User Manual

Page 11

Advertising
background image

9

Digital (Enhanced) Overhead Stirrer

Do not remove the power line when cleaning. Only use
recommended cleansers:

Dyes

Isopropyl alcohol

Construction materials Water containing tenside / Isopropyl

alcohol

Cosmetics

Water containing tenside / Isopropyl

alcohol

Foodstuffs

Water containing tenside

Fuels

Water containing tenside

Before using other method for cleaning or decontamination,

the user must ascertain with the manufacturer that this

method will not damage the instrument. Wear the proper

protective gloves during cleaning of the instrument.

Note:

Electronic device can not clean with cleanser.

If you require maintenance service, must be

cleaned the instrument in advance to avoid

pollution of hazardous substances, and to

send back into original packing.

If the instrument will not use for a long

time,please switch off and place in a dry,

clean, room temperature and stable location.

9.Interface and output

Overhead stirrer makes use of special accessories and cables

to connect external devices. Standard 9-pin interface is used

to be connected to the PC.

selected from the specified communication line of the
EIA-standard RS232C, corresponding with DIN66020.

Transmission method: Asynchronous signal transmission
in start-stop-operation.

Mode of transmission: Fully Duplex. 1 starts bit; 7
character bits; 1 stop bit.

Transmission speed: 9600 bit/s.

Data communication from laboratory instrument to
computer is only possible on demand of the computer.

Note:

D o n o t i n s e r t o r r e m o v e t h e R S 2 3 2 C

communication line when switch on!

10.Maintenance and Cleaning

Proper maintenance can keep instrument working
properly and lengthen its lifetime.

Do not spray cleanser into the instrument when cleaning.