Safety instructions – Scilogex RE100-Pro User Manual

Page 4

Advertising
background image

2

Rotary Evaporator

1. Safety Instructions

Read the operating instructions in full before starting

up and follow the safety instructions.

Keep the operating instructions in a place where they

can be accessed by everyone.

Ensure that only trained staff work with the appliance.

Follow the safety instructions, guidelines, occupational

health and safety and accident prevention regulations.

When working under a vacuum in particular!

Wear your personal protective equipment in

accordance with the hazard category of the medium to

be processed. Otherwise there is a risk of:

- splashing liquids,

- body parts, hair, clothing and jewellery getting

caught,

- injury as a result of glass breakage.

CAUTION! Inhalation of or contact with media such

as poisonous liquids, gases, spray mist, vapours, dusts

or biological and microbiological materials can be

hazardous to user.

Set up the device in a spacious area on an even, stable,

clean, non-slip, dry and fireproof surface.

Ensure that there is sufficient space above the device

as the glass assembly may exceed the height of the

device.

Prior to each use, always check the device, accessories

and especially the glass parts for damage. Do not use

damaged components.

Ensure that the glass assembly is tension-free! Danger

of cracking as a result of:

- stress due to incorrect assembly,

- external mechanical hazards,

- local temperature peaks.

Ensure that the stand does not start to move due to

vibrations respectively unbelance.

Beware of hazards due to:

- flammable materials,

- combustible media with a low boiling temperature,

- glass breakage.

CAUTION! Only process and heat up media that has a

flash point higher than the adjusted safe temperature

limit of the heating bath that has been set.

The safe temperature limit of the heating bath must

always be set to at least 25 °C lower than the fire point

of the media used.

Advertising