Supported load, Faults, Interface – Scilogex SK-O180-Pro User Manual

Page 10

Advertising
background image

8

SK-O(L)330- Pro Orbital and Linear Digital Shaker SK-O(L)180- Pro Orbital and Linear Digital Shaker

the mains switch. After about 5 second, you can let go of the

knobs.

• Switch between operating Modes A and B in order.

6.4 Switch the shaking direction

Change the shaking direction according to the steps as

follows:

• Turn OFF the instrument;

• Press and hold time knob, at the same time turn on the

instrument, release the time knob after 10 seconds;

• Switch the operation direction in turn.

7.Supported load

In order to ensure safe, the shaker must only be operated

within the supported weight (7.5kg for 330 and 2.5Kg for

180).

Make sure the holding surface of the shaker is always clean

and level. Ensure the individual shaking vessels must be

placed evenly and fastened securely in the middle of the

shaking table no matter one or multiple shaking vessels are

placed.

8.Faults

• When switched ON, the instrument doesn’t work

Check whether the power cable fitted well

Check whether the fuse is broken

• The speed cannot reach set value

Check whether it is overloaded

• The motor does not start via pressing the speed knob and

time knob

Check whether the time is set to zero

9.Interface

SK-O330(180)-Pro/ SK-L330(180)-Pro Orbital and Linear

Digital Shaker, special adapter and cable are used to link to

external device.and the Standard 9-pin interface to connect

with the PC.

• The cable between the shaker and the computer used EIA

Standard RS232C communication line,corresponding to

the DIN 66020 interface.

• Transmission method: Asynchronous signal transmission.

• Mode of transmission: Fully Duplex.

• 1 start bit; 8 character bits; 1 stop bit.

• Transmission speed: 9600 bit/s

Note:

Forbid to plug in and plug off the power cable

when the instrument working