Fill and vent circuits – Taylor-Wharton EF-450 User Manual

Page 8

Advertising
background image

Fill and Vent Circuits


The liquid valve (V-1) communicates with the bottom of the vessel. A stainless steel tag labeled “LIQUID”
identifies the valve and the liquid connection (CN-1). Liquid is added or removed from the vessel through
this connection and valve.

The vent / trycock valve (V-2) is attached to a vertical tube in the upper portion of the vessel. The open end of
the tube is positioned at 90% liquid level based on the vessel volume. Opening the vent valve reduces
pressure in the vessel during filling. It also severs as a “full trycock”, venting liquid from the vessel when the
liquid level exceeds 90%. A tag labeled “VENT” is attached to this valve.

V-1

CN-1

V-2

CN-2

Figure 2: Fill and vent circuits highlighted in blue.

8