Salter Brecknell 650 User Manual

Page 39

Advertising
background image

39

“P”

- This command is sent to the scale to print the indicated display. The

scale will not respond if the scale is in motion, positive overload or

negative overload.

“Z”

- This command is sent to the scale to zero the scale. If the scale is in

motion, the scale will wait until a state of equilibrium is achieved. The

scale will not respond at all if the scale is in positive overload or

negative overload. The scale will also not respond if it is not in gross

mode or within the zero range specified in F4 of the Setup Menu.

“T”

- This command is sent to the scale to tare the scale. If the scale is in

motion, the scale will wait until a state of equilibrium is achieved. The

scale will not respond at all if the scale is in positive overload or

negative overload. The scale will also not respond if it displaying a

negative gross value.

“G”

- This command is sent to the scale to revert to gross mode. The scale

will not respond if the scale is in motion, positive overload or negative

overload. The scale will also not respond if it is not in net mode.

“N”

- This command is sent to the scale to revert to net. The scale will not

respond if the scale is in motion, positive overload or negative

overload. The scale will also not respond if it is not in gross mode or a

tare has yet to be established.

“C”

- This command is sent to the scale to toggle among the configured

units.

TABLE B-1. Recognized Host Commands

B.2.2.2

CONTINUOUS MODE

The Demand mode is used to interface to computers, scoreboards and other remote

devices requiring constant data updating. The transmission occurs at the end of

each display update. Figure B-9 shows the serial data format for the Continuous

Mode.

<STX> <POL> xxxxx.xx

<L/K> <G/N> <STAT> <CR> <LF>

Polarity:

<SP> = Positive

"–" = Negative

Weight Data

Units:

L = pound

K = kilogram

Gross/Net:

G = Gross

N = Net

Start

Transmission

Carriage

Return

Line

Feed

Status:

<SP> = Valid

M = Motion

O = Over/under range

P = pieces

FIGURE B-9. Consolidated Controls Continuous Mode

Advertising