Appendix b: 6505 shielding effectiveness test kit – ETS-Lindgren 6512 Shielded Loop Antenna User Manual

Page 37

Advertising
background image

www.ets-lindgren.com

6505 Shielding Effectiveness Test Kit

37

Appendix B: 6505 Shielding Effectiveness

Test Kit

The ETS-Lindgren Model 6505 Loop Shielding Effectiveness Test Kit is a set of
loop antennas that includes the following:

Model 6507 Active Receiving Loop Antenna

Model 6509 Passive Loop Antenna

Battery Charger

Measuring Magnetic Shielding Effectiveness

1. Set the transmitting and receiving antennas apart at a distance equal

to 24 inches plus the thickness of screen room.

2. Set up equipment as shown in the diagram on page 39.

3. Turn generator to on position.

4. Disable RF output of generator.

5. Turn amplifier to on position.

6. Turn receiver to on position.

7. Turn receiver loop power on, and verify light is on.

8. Set desired frequency on generator. Set the same frequency on

receiver.

9. Set the attenuator at maximum (100–120 dB) attenuation.

10. Enable RF output of the generator.

Advertising