ETS-Lindgren HI-2790B Calibration Comparison System User Manual

Page 20

Advertising
background image

20

Operation

www.ets-lindgren.com

4.1.10

Check the system field level at 1 mW/cm

2

by placing the

HI-1710 LCR probe in the chamber and aligning the probe at its

midpoint reading.

4.1.11

Remove the HI-1710 LCR probe and place it in its holder on top of

the system.

4.1.12

This ends the setup procedure for the HI-2790B system. The

information logged in step 4.1.9 on page 19 must be reviewed to

make sure that the HI-2790B system remains stable. No value of

the DIF or the RF Power reading may be lower than 10% less than

the highest reading observed over a 12-month period between

annual calibrations of the HI-1710 LCR. If a reading falls below

these limits, carefully review and repeat the setup procedure. If it is

still outside the limits, check the rest of the data for indications of

the problem area. The HI-1710 LCR bias, ellipticity, and the

ambient temperature readings should also remain within specified

limits.