against, and have passed the applicable

requirements of relevant electromedical

standards, including EN 60601-1-2:2007.

The pro5™ and pro5 AIRdaptive™ models

must emit electromagnetic energy in

order to perform their intended function.

Nearby electronic equipment may be

affected by this emission. Similarly

the pro5™ and pro5 AIRdaptive™

models may themselves be affected by

electromagnetic emissions from other

equipment in the vicinity.

The pro5™ and pro5 AIRdaptive™

models are intended for use in the

electromagnetic environment specified

below. The customer or user of the

pro5™ or pro5 AIRdaptive™ model should

assure that they are used in such an

environment:

– Floors should be wood, concrete or

ceramic tile. If the floor is covered with

synthetic material, the relative humidity

in the room should be at least 30%.

– Mains power quality should be that

of a typical domestic, commercial or

hospital environment as appropriate.

– The pro5™ and pro5 AIRdaptive™

models should not be used during

power mains interruptions

– Power frequency magnetic fields

should be at levels characteristic of a

typical location in a typical commercial

or hospital environment.

The pro5™ and pro5 AIRdaptive™

models are intended for use in an

electronic environment in which radiated

radio-frequency (RF) disturbances

are controlled.

The customer or user of the pro5™

or pro5 AIRdaptive™ model can help

prevent electromagnetic interference by

maintaining a minimum distance between

portable and mobile RF communications

equipment (transmitters) and the

pro5™ or pro5 AIRdaptive™ model

as recommended below, according

to the maximum output power of the

communications equipment:

The adequate separation distance

to provide base immunity to RF

disturbances is 1.0m.

Retain this user manual for your future

reference.