0 troubleshooting and testing, 0 maintenance, 1 basic maintenance – Rice Lake RL550-10 Digital Chair Scale Operating Instructions User Manual

Page 20: 2 cleaning

Advertising
background image

16

Digital Chair Scale Operation Manual

6.0

Troubleshooting and Testing

Refer to the following instructions to check and correct any failure before contacting service personnel.

Table 6-1. Troubleshooting Table for the Rice Lake Scale Line

Symptom

Possible Cause

Corrective Action

Scale does not turn on

Dead battery

Connect the scale to a power source.

Faulty electrical outlet

Use a different electrical outlet.

Bad power supply

Replace optional adapter.

Questionable weight or the scale
does not zero

External object is interfering with the
scale

Remove the interfering object from the scale.

Display did not show 0.0 before
weighing

Help the patient off the scale, zero the scale and
begin the weighing process again.

Scale is not placed on a level floor

Ensure the scale is level and begin the weighing
process again.

Scale is out of calibration

Check the weight with a known weight value.

The display shows a STOP message

The load on the scale exceeds the
capacity of the scale

Remove the excess weight and use the scale
according to manufacture’s specs.

The display shows LO Bat message

The battery is low

Recharge the battery.

The display shows Err message as
detailed in the table below

Err 2

Low saturation state (low A/D)

The load cell is not connected properly. Check
the cables and mechanical connections. If the
problem persists, replace the set of load cells.

Err 3

High saturation state (high A/D)

See Err 2

Err 6

Unstable weight. Cannot calibrate

Check the load cells’ mechanical surroundings
and see that nothing touches them and that the
cables are properly welded.

7.0

Maintenance

The following section provides instructions for maintaining and cleaning the Rice Lake line of scales.

Maintenance operations other than those described in this section should be performed by qualified service

personnel.

7.1

Basic Maintenance

Before the first use of the scale and after periods of non-use, check the scale for proper operation and function. If

the scale does not operate correctly, contact qualified service personnel.
Go through the following steps for basic maintenance.

1. Check the overall appearance of the entire scale for any obvious signs of damage, abuse, etc.
2. Inspect the condition of the optional AC adapter for cord cracking or fraying or for broken or bent prongs.

7.2

Cleaning

Proper care and cleaning is essential to ensure a long life of accurate and effective operation. Before beginning

the cleaning process, disconnect the scale from the AC power source.

1. Clean all external surfaces with a clean, damp cloth or tissue. Mild soap and water solution may be used.

Dry with a clean soft cloth.

2. Do not immerse the scale into cleaning or other liquid solutions.
3. Do not use Isopropyl alcohol or other solutions to clean the display surface.

Advertising
This manual is related to the following products: