Analog output calibration – Rice Lake 920i Programmable HMI Indicator/Controller - Analog Output Card User Manual

Page 3

Advertising
background image

Analog Output Card Installation

3

Analog Output Calibration

The following calibration procedure requires a multimeter to measure voltage or current output from the analog
output module.

Note

The analog output must be calibrated after the indicator itself has been configured and calibrated.

1. Enter setup mode and go to the ALGOUT menu (see Figure 3):

• Set SOURCE1 for the number of the scale channel used as the source for the analog output

• Set MODE1 to track either gross or net weight from that scale

• Set OFFSET to 0% for 0–10 V or 0–20 mA output; set to 20% for 4–20 mA output

• Set ERRACT to specify how the analog output will respond to system error conditions

• Set MIN to lowest weight value to be tracked by the analog output

• Set MAX to highest weight value to be tracked by the analog output

• For the 880 and 820i, MIN NEG and MAX NEG parameters allow tracking of negative weight values.

•MIN NEG - set to on to specify the minimum weight (MIN parameter) is a negative value.

•MAX NEG - set to on to specify the maximum weight (MAX parameter) is a negative value.

SCALES

SERIAL

FEATURE

PFORMT

SETPTS

DIG I/O

ALGOUT

VERS

number

40

number

SOURCE

PROG

OFFSET

20%

0%

ERRACT

SCALE n

FULLSC

HOLD

ZEROSC

TWZERO

59650

TWSPAN

MODE

GROSS

NET

If SOURCE1

PROG

n = slot number of analog output card

ALGOUTn

ALGOUTn+14

If dual-channel card

installed

number

10000

ON

MIN

number

000000

MINNEG

ON

OFF

MAX

OFF

MAXNEG

Figure 3. Analog Output Menu

2. Connect multimeter to connector J1 on the analog output card:

• For voltage output, connect voltmeter leads to pins 3 and 4
• For current output, connect ammeter leads to pins 1 and 2

3. Adjust zero calibration: Scroll to the TWZERO parameter. Press to view zero value, then check voltage

or current reading on multimeter. Use the numeric keypad to adjust the zero value up or down.

4. Adjust span calibration: Scroll to the TWSPAN parameter. Press to view span value, then check voltage

or current reading on multimeter. Use the numeric keypad to adjust the span value up or down.

5. Final zero calibration: Return to the TWZERO parameter and verify that the zero calibration has not

drifted. Use the numeric keypad to adjust the zero value as required.

6. Return to normal mode. Analog output function can be verified using test weights.

For dual-channel cards, repeat calibration steps for the second (J2) connector.

See the 920i Installation Manual (PN 67887) or 820i Installation Manual (PN 93018) for more configuration
information.