Cable Grounding

All cables routed into the panel mount enclosure
should be grounded against the indicator enclosure. Do
the following to ground shielded cables:

Use the lockwashers, clamps, and kep nuts
provided in the parts kit to install grounding
clamps on the enclosure studs adjacent to cord
grips.

Figure 4. Grounding Clamp Attachment for Foil-Shielded

and Braided Cabling

Figure 5. Cable Installation with Grounding Clamp

Route cables under cabling bar (see Figure 5) and
through grounding clamps to determine cable
lengths required to reach cable connectors. Mark
cables to remove insulation and shield as described
below:

For cables with foil shielding, strip
insulation and foil from the cable half an
inch (15 mm) past the grounding clamp
(see Figure 4). Fold the foil shield back on
the cable where the cable passes through
the clamp. Ensure silver (conductive) side
of foil is turned outward for contact with
the grounding clamp.

For cables with braided shielding, strip
cable insulation and braided shield from a
point just past the grounding clamp. Strip
another half inch (15 mm) of insulation
only to expose the braid where the cable
passes through the clamp.

For load cell cables, cut the shield wire just
past the grounding clamp. Shield wire
function is provided by contact between
the cable shield and the grounding clamp.

Route stripped cables under cabling bar
and through the clamps. Ensure shields
c o n t a c t g r o u n d i n g c l a m p s . Ti g h t e n
grounding clamp nuts.

Use cable ties to secure cables to cabling
bar.

Power Specifications

Line Voltages

115 or 230 VAC

Frequency

50 or 60 Hz

Power Consumption

25W maximum

See the

820i

Installation Manual for additional

specifications.

Parts Kit Contents

Table 1 lists the parts kit contents for the panel mount
version of the

820i

.

Insulated cable

Foil (silver side out)

Grounding clamp

Shield wire (cut)

Length of foil before folding

back on cable insulation

Cut insulation here

for foil-shielded cables

Braid

Cut insulation here

for braided cables

Install lockwashers first,

against enclosure, under

grounding clamp

BRAIDED CABLE

FOIL-SHIELDED

CABLE

Grounding clamp

Install lockwashers first,

against enclosure, under

grounding clamp

Cabling bar

Use cable ties to

bind cables to bar

Grounding clamp stud

PN

Description

14626

Kep nuts, 8-32NC (5)

54206

Machine screws, 6-32 x 3/8 (2)

15133

Lock washers, No. 8, Type A (5)

71522

Machine screws, 8-32NC x 1/4 (6)

82425

Machine screws, 10-32NF x 1-1/2 (9)

15631

Cable ties (4)

53075

Cable shield ground clamps (5)

77925

Revolution

™ CD (1)

15887

6-position screw terminal for load cell
connection (1 or 2)

76514

6-position screw terminal for J10 (1)

76513

4-position screw terminal for J9 (1)

71344

10-position screw terminals for J5 and J11 (2)

Table 1.Parts Kit Contents