Calibration

5

3.0

Calibration

It is recommended that the scale be “exercised” by loading it to near capacity two or three times before calibration

to be certain that everything is seated. To calibrate the Benchmark HD scale place test weights on the scale

platform equal to 70 - 100% of the scale’s capacity. If several weights are used, distribute them evenly around the

platform.
See your indicator manual for specific indicator calibration procedures.