RL-DBS Operating Instructions - Troubleshooting & Testing

5

5.0

Troubleshooting & Testing

Refer to the following instructions to check and correct any failure before consulting service personnel.

Symptom

Possible Cause

Corrective Action

Scale does not turn on when using just
batteries

Dead batteries

Connect the scale to a power source or
change batteries

Scale does not turn on

Dead battery

Connect the scale to a power source

Faulty electrical outlet

Use a different electrical outlet

Bad power supply

Replace adaptor

Questionable weight or the scale does
not zero

External object is interfering with the scale Remove the infant/interfering object from

the weighing tray from the scale

Display did not show 0.0 before weighing

Remove the infant, zero the scale and
begin the weighing process again

Scale is not placed on a level surface

Place the scale on a stable surface and
begin the weighing process again

The weighing tray is not placed properly

Place the weighing tray in its proper place

Scale is out of calibration

Check the scale with a known weight
value

The display shows o_Ld message

The load on the scale exceeds the
capacity

Remove the excess weight and use the
scale according to its stated limit

The display show Err

The RECALL key was pressed with
insufficient weight on the scale.

Place more than 2 lb, 30 Oz on the scale

Table 5-1. Troubleshooting Table