2 communication protocols, 1 escape protocol – Rice Lake Dual Ramp Wheelchair Platform Scale with Seat (350-10-4) - Rice Lake User Manual

Page 23

Advertising
background image

Wheelchair Scale with Seat Technical Manual - RS-232 Communication

19

7.2

Communication Protocols

The Rice Lake Digital Wheelchair Scale with seat has two communication protocols, escape and maintenance protocol.

7.2.1

Escape Protocol

An escape protocol is where the escape (0X1B or ASCII 27) is used to indicate that there is a command

following. On the PC side there must be a listener created by the vendor that will interpret this protocol. This

listener must also take care of all the issues regarding data integrity, etc. to make sure that the data that was sent

and received is valid.
Two examples include:

Scale initiated communication

PC initiated communication

Table 7-1 is what can be sent across communications lines.

Table 7-1. Escape Protocol Commands

PC Initiated

ESC Value

Request current values/settings

R

Diagnostics

A

Send scale control messages

C

PC Initiated

ESC Value

Send single reading

R

Send diagnostic response

Table 7-2 lists the ESC characters that will be used.

Table 7-2. ESC Characters

Name

ESC character

ESC value with

parameters

Description

Reading

R

R

Tells the PC that the scale is sending a reading. Immediately following
this is the value that is sent. Example:
<ESC><R>ESC><W0200.0<ESC>Nm<ESC>E

Weight

W

Wnnn.n

Is the patient weight (example: W02000 means 200.0). If the scale is
overloaded or under loaded, the scale will return the value 999.99.

Height

H

Hnnn.n

Patient height

BMI

B

Bnn.n

Patient BMI

Units

N

Nc

Indicates in which unit the values have been taken (m=metric,
c=constitutional).

End of Packet
(EOP)

E

E

Indicates that the end of the command has been reached.

Diagnostics
(request)

A

Accc

A request for a diagnostic test on certain parts of the scale (such as
battery life, load cells).

Diagnostics
(response)

Z

Zccc

This will be the response of the diagnostics done on the scale. Values
will include error codes to indicate what is wrong with the scale, or all
zeros (Z000) to indicate that all is well.

Control (set a
value)

C

Cccc=c

Sets the value of the scale’s global settings. Example:
<ESC><CUOM=m><ESC><E will set the unit of measurement.

Table 7-3. Scale Global Values List and Identifiers

Name of Control

Identifier

Unit

Unit of Measure (metric or constitutional)

UOM

c (m or c)