Operating instruction, Error messages – Rice Lake DWM-IV Weighmeter Series 9 Operation Manual User Manual
Page 3
FlexWeigh
q
Series 9 Operation Manual • Rev 2.00
Page 3 of 4
Operating Instruction
q
Power On
With nothing on the platform, press the key . The display unit wil l turn on.
q
Zero
To zero the scale, press t he key.
q
Tare by Push but ton
Place the object to tare on the scale and press the key.
Error Messages
Error Messages Explanation
CAP OL
Over Capacity, weight value exceeds the Full Scale Capacit y
ADC UL
Input load cell signal is too low or reversed, load cell or load cell-cable is faulty
ADC OL
Input load cell signal is too high, load cell or load cell-cable is faulty
NOT CAL
Scale is Not yet Calibrated. Need to calibrate the scale
RMT DSP
Displays on Secondary Display when no signal from Primary Display
SPAN OL
Max calculated span count is greater than maximum ADC Chip count.
LO BAT
Insufficient battery voltage level. Need to re-charge the battery
- - - - - - -
Displays if Power On Zero fails. Need to unload the scale or press the Zero Key
if power on zero weight is within allowable zero range