8

Rice Lake Digital Floor Level Scale Operation Instructions

5.0

RS-232 Communication

The scale comes with an RS-232 port that enables weight data to be transmitted to other equipment, such as a

computer or printer. The RS-232 cable with DB-9 connector (PN 100719) is available from Rice Lake Weighing

Systems. Figure 5-1 shows where the RS-232 connection is.
The RS-232 parameters are 9600 baud (selectable in the programming mode), 8 data bits, 1 stop bit, no parity

and no handshaking.
There are three methods of communication:

Pushbutton keypad print

Standard remote protocol

Escape protocol

5.1

Pushbutton Keypad Print

With a stable, in-range weight, press and hold the

LB/KG/Print

key for at least three seconds, or until the scale

displays

PRINT

. Note that if the scale does not show print, release the button as the weight was either in motion or

out of range.

If displaying weight and not BMI, the scale will send out the following 21 character string:

xxxxxxxxx<SP>uu<SP>mmmmm<SP><CR><LF>
Where:

xxxxxxxxx is the weight with decimal point and " - " sign, if negative uu is the unit (LB or KG).
mmmmm is the mode (gross or net)

Examples:

-10 Lb net = <SP><SP><SP><SP>-10.0<SP>lb<SP><SP>Net<SP><SP><SP><CR><LF>
10 Lb gross = <SP><SP><SP><SP><SP>-10.0<SP>lb<SP>Gross<SP><CR><LF>

In BMI mode (displaying the BMI value), the scale will send out the following data:

GROSS WEIGHT 215.0 LB
TARE WEIGHT 0.0 LB
NET WEIGHT 215.0 LB
PATIENT HEIGHT 6-01.0 FT
PATIENT BMI 28.4

5.2

USB Connection

The Rice Lake Digital Floor Level Scale has the capability of connecting to a PC using a USB connection and a

USB cable (not included). That connection location is shown in Figure 5-1.

USB Connection

RS-232 Connection

Figure 5-1. USB Connection Port and RS-232 Connection Port

Connecting software and downloads should be addressed by your IT department. For more information on

setting up the USB connection, please refer to the RL150-10-7 Digital Floor Level Scale Technical and

Operating Instructions, PN132098, which is available online at www.ricelake.com/health.