3.0

Calibration

Power up the indicator and allow the load cell to warm up for 20 minutes before calibrating.

1. With the deck in place and no additional weights on the scale, set zero on the digital indicator according to

the procedure described in the indicator’s operating manual.

2. Place certified test weights on the scale, using 75-100% of the scale capacity.
3. Span the meter to actual capacity using the procedure in the operating manual.
4. Remove the test weights and re-zero the indicator.
5. If required, repeat the above procedure until the unit is correctly calibrated.

Important