0 appendix, 1 troubleshooting, Table 4-1. troubleshooting – Rice Lake Floor Scales - SURVIVOR FB Series Electronic Flexure Base User Manual

Page 9: 2 load cell replacement, 1 troubleshooting 4.2 load cell replacement

Advertising
background image

Appendix

5

4.0

Appendix

4.1

Troubleshooting

4.2

Load Cell Replacement

1. Remove the deck assembly (see section 2.1 on page 2).
2. Loosen the lock nut on the shock-stop bolts shown in Figure 4-1 and turn them down against the angle

until finger-tight. This locks in the correct lever height for the load cell.

Problem

Possible Cause

Zero shift

• Improper shimming under base corners.
• Loose fastener.
• Crooked flexures.
• Obstruction under live section.
• Damaged load cell or indicator.

Unstable readings

• Improper shimming under base corners.
• Load cell cable ground-shield not connected.
• Moisture in load cell or signal cable.
• Crooked flexures.
• Obstruction under live section.
• Indicator configured inappropriately.
• Damaged load cell or indicator.

Different corner readings

• Improper shimming under base corners.
• Loose fastener.
• Crooked flexures.
• Obstruction under live section.
• Corner correction shimming required.

Non-linearity

• Primary lever too high or too low.
• Central flexure misalignment.
• Crooked flexures.
• Obstruction under live section.
• Damaged load cell or indicator.

Readings differ when same

weight is placed on different

sides of deck

• Central flexure misalignment.
• Improper shimming under base corners.
• Primary lever too high or too low.
• Obstruction under live section.
• Corner correction shimming required.

Table 4-1. Troubleshooting