IQ plus 350 Analog Output Module Installation

To install or replace the analog output module, ensure
power to the indicator is disconnected, then place the
indicator face-down on an antistatic work mat.
Remove the screws that hold the backplate to the
enclosure body. Loosen all in-use cord grips then lift
the backplate away from the enclosure and set it aside.

Use a wrist strap to ground yourself and
protect components from electrostatic
discharge (ESD) when working inside
the indicator enclosure.

NOTE:

Ensure jumpers JP1 and JP2 are installed in

position 1 (POS1) as shown in Figure 7.

Mount the analog output module on its standoffs in
the location shown in Figure 7 and plug the module
input into connector J5 on the IQ plus 350 board.
Connect output cable to the analog output module as
shown in Table 3, then reassemble the enclosure as
described on page 6.

Table 3. Analog Output Module Pin Assignments

Figure 7. Analog Output Module Installation and Wiring on IQ plus 350 Board

Caution

Pin

Signal

1

+ Current Out

2

– Current Out

3

+ Voltage Out

4

– Voltage Out

SW1

J7

JP4

J1

J4

F1

LOAD CELL CONNECTOR

SERIAL COMM & DIGITAL IN

KEYPAD CONNECTOR

F2

JP7

Setup Switch

Serial Communications Cable
to J4 Connector

Load Cell Cable
to J1 Connector

Line Filter Load Wires

to JP7 Connector

Setup Switch Wires

to JP4 Connector

Cable tie mounts on inside of enclosure top

Cable ties

JP1

JP2

Analog Output

Cable

7

6

5

4

3

2

1

1

2

3

4

5

6

7

J5

ANALOG OUTPUT

ANALOG OUTPUT MODULE

J1

4

3

2

1

JP1

JP2

POS1

POS2

POS2

POS1

Advertising
This manual is related to the following products: