Calibration

23

4.2

EDP Command Calibration

To calibrate the indicator using EDP commands, the
indicator EDP port must be connected to a terminal or
personal computer. See Section 2.3.3 on page 5 for
EDP port pin assignments; see Section 5.0 on page 24
for more information about using EDP commands.

Once the indicator is connected to the sending device,
do the following:

1. Place the indicator in setup mode (display

reads

CONFIG

) and remove all weight from

the scale platform. If your test weights require
hooks or chains, place the hooks or chains on
the scale for zero calibration.

2. Send the WZERO EDP command to calibrate

zero. The indicator displays

*CAL*

while

calibration is in progress.

3. Place test weights on the scale and use the

WVAL command to enter the test weight
value in the following format:

WVAL=nnnnnn<CR>

4. Send the WSPAN EDP command to calibrate

span. The indicator displays

*CAL*

while

calibration is in progress.

5. To remove an offset value, clear all weight

from the scale, including hooks or chains used
to hang test weights, then send the REZERO
EDP command. The indicator displays

*CAL*

while the zero and span calibrations are
adjusted.

6. Send the KUPARROW EDP command to exit

setup mode.

4.3

Revolution

Calibration

To calibrate the indicator using Revolution, the
indicator EDP port must be connected to a PC running
the Revolution configuration utility.

Use the following procedure to calibrate the indicator:

1. Select

Calibrate Indicator

from the Revolution

main menu.

2. On the Indicator Calibration display, select

the indicator model (

IQ+390-DC

) and

communications port, then click

OK

.

3. Revolution uploads calibration data from the

indicator then presents the information in a
display like that shown in Figure 4-3.

Figure 4-3. Revolution Calibration Display

4. Enter the

Value of Test Weight

to be used for

span calibration then click

OK

.

5. The Zero Calibration dialog box prompts you

to remove all weight from the scale. Clear the
scale and click

OK

to begin zero calibration.

NOTE:

If your test weights require hooks or

chains, place the hooks or chains on the scale
for zero calibration.

6. When zero calibration is complete, the Span

Calibration dialog box prompts you to place
test weights on the scale for span calibration.
Place tests weights on the scale then click

OK

.

7. When span calibration is complete, the

Rezero dialog box prompts you to remove
weights from the scale. Remove the weights
then click

OK

.

8. When calibration is complete, the

New

Settings

fields of the Indicator Calibration

display are filled in. Click

Exit

to save the new

values and return to the Revolution main
menu; to restore the previous calibration
values, click

Restore Settings

.