Testing methods, Unlocking methods – Rice Lake MA160 Sartorius Moisture Analyzer User Manual

Page 48

Advertising
background image

48

Moisture Analysis

Operating Instructions Moisture Analyzer MA160

Testing Methods

Evaluation

Result is within the tolerance range
Result: 85.15 %M
Target value: 85.00 %M
Tolerance range: 84.50 %M - 85.50 %M
Are you satisfied with this result?

NO

YES

y If the target value and initial weight options are activated, the measurement/

method is evaluated and the result of the evaluation is displayed .

t

Select the desired option:

– If you are satisfied with the result, select

YES

. You can then unlock the

method .

– If you are not satisfied with the result: Select

NO

and run the test again .

Test the method

Method successfully created. Would you

like to test the method now?

Temperature: 105 °C

NO

YES

t

Select the desired option:

– To test the method with the suggested temperature: Select

YES

. Moisture

analysis is then carried out with the set parameters (see „Performing the
Moisture Analysis“ on page 38)
.

– To save the method without testing: Select

NO

and then

o

to return to the

method menu .

Unlocking Methods

+

Standard

Coffee creamer

100 °C
Fully automatic

Oats

80 °C
Fully automatic

Wood glue

110 °C
Fully automatic

105 °C
Fully automatic

Ketchup

110 °C
Fully automatic

Methods must be unlocked (available) before they can be used .

Unavailable (locked) methods are marked in the method menu with an

H

symbol .

t

Select the method you wish to unlock .

Customize parameters

Heating program

End

Initial weight

Result display

Standard drying
110 °C

Fully automatic

Off

Moisture in %M

Method name

Ketchup

y The parameters of the method are displayed .

Customize parameters

Test

Release

Delete

Print

t

Scroll all the way to the bottom .

t

Select

Release

.

Advertising