The following procedure MUST be performed to properly save and exit from the
MSI-9000 CellScale setup / configuration menus when used in conjunction with a
MSI-6260CS and operated by version 5-20, or later MSI-9000 application software.


This procedure is to be performed after ALL calibration(s) and configuration(s) are complete. It also
MUST be performed after any connection to a PC, even if no changes were made.

1). Perform all calibration & setup as required per application and exit to Main Menu.

Select (1) Channels / Calibration.
















2). Select (2) Data backup.
















3). Select (b) Backup data NOW!









---------- MSI 9000 - CellScale ----------
- (c) 2006 Measurement Systems International -
----------------------------------------------

16:57:52 01/05/08

(1) Channels/Calibration

(2) Communications
(3) Display
(4) Feature Locks/Passwords
(5) Product IDs
(6) Set Points
(7) System
(8) About...

Enter choice :

***** Channel Setup *****

(1) Channel Calibration

(2) Data backup

(3) Filter (software)
(4) Channel Scanlist functions
(5) Motion Compensation
(6) Motion Detection
(7) Tare
(8) Total
(9) Units Lock
(a) Zero
(b) Standard Mode: Industry
(c) 4 to 20mA
(d) Temperature Sensor Installed: No
(ESC) Exit

***** Data Backup *****

(1) Backup data after every cal: Yes
(2) Restore from ROM every power up: No
(3) Restore from ROM including Slave and Math loadcells: Enabled

(b) Backup data NOW!

(r) Retrieve backup data NOW!
(ESC) Exit