The MSI-9000 CellScale calibrates with the first test weight, and then presents the
following menu. The upper half of the screen indicates the two cal steps taken, zero and
a 10000 kg test weight. Press ESC to end calibration, and store all the calibration
constants.













12)

Calibration is now complete. NOTE: It may take up to 30 sec. to complete store & save
process. When complete, Main Menu should appear. If after 30 sec. it does not appear in
full, press ENTER.
















13)

It may be required to set the “UNITS” of the 9300 to indicate KG. Refer to full version
MSI-9300 User’s Guide for details.

If 5-20, or later, MSI-9000 CellScale application software is used. Perform

MSI-9300-5-20 Setup Exit

procedure before exiting system.



If software version is less than 5-20, then disconnect comm. cable and

cycle power on scale.


.


***** Channel Calibration *****

Calibrating Ch # 1-1, named:

.................

Precision in Kilograms :

10000. x 2.

Test weight # 0 in Kilograms :

0.

Test weight # 1 in Kilograms :

10000.



Test Weight # 2 in Kilograms:

10000

Press (Y)

when scale is loaded and steady

Press (N)

to

change

Press (B)

to backup

Press (ESC) to complete and store calibration

---------- MSI 9000 - CellScale ------------------------
-(c) 2005 Measurement Systems International -
----------------------------------------------------------------
13:48:06

01/01/08


(1)

Channels/Calibration

(2)

Communications

(3)

Display

(4)

Feature Locks/Passwords

(5)

Product IDs

(6)

Set Points

(7)

System

(8)

About...


Enter choice: